In the literature a good reliability of isokinetic measurements is rep
orted. In spite of these reports high variations of the peak torque an
d work values were registered when testing 24 patients with low back p
ain, using a CYBEX TR isokinetic trunk rotation apparatus. After a 3 w
eek's training the patients force (peak torque) increased in the avera
ge between 10.7 p.c. (SD+/-26.4) and 23.2 p.c. (SD+/-63.8), correspond
ing to the test-velocity. Seven of the 24 patients improved or deterio
rated more than 30 p.c. Analysing the parameters which could interfere
with the results there was found a discrepancy between the original C
YBEX software and the German Iso-Kin 4.0 software. The Isokin 4.0 does
not warm up and calibrate the isokinetic apparatus daily. Investigati
ons with definite weights showed that the values for peak torque are 4
0-50 m higher when using the Iso-Kin 4.0 software in comparison to the
original CYBEX software. Furthermore there are a lot of other paramet
ers such as acceleration of the body, psychological alterations, pain,
motivation, learning effects, and artificial oscillations of the regi
strated curve, which can affect the results even when using the origin
al CYBEX software.