I. Penttila et al., ADAPTATION OF THE TROPONIN-T ELISA TEST TO A MICROPLATE IMMUNOASSAY READER, European journal of clinical chemistry and clinical biochemistry, 33(1), 1995, pp. 59-63
Troponin T concentration in serum is usually measured by the automated
method developed by Boehringer Mannheim for the ES-series of analyser
s. These instruments need at least 140 mu l of serum and 700 mu l of r
eagents for a single analysis, which takes 90 min. We describe an alte
rnative procedure, using streptavidin-coated microtitre plates, tropon
in T reagents of Boehringer Mannheim and an ELISA-reader to measure th
e concentration of troponin T. The present assay needs only 30 mu l of
sample and 200 mu l of reagents, and it takes 75 min; the detection l
imit is 0.10 mu g/l. We also assessed the microtitre plate method for
sensitivity and precision and compared the results with those measured
by an ES-300 automatic analyser. Both methods have the same measureme
nt range for troponin T of 0.1 to 15 mu g/l. For daily routine use of
the microtitre plate method we recommend duplicate determinations.