THE BOUNDARY INDUCED ERROR ON THE MEASUREMENT OF THERMAL-CONDUCTIVITYBY TRANSIENT HOT-WIRE METHOD

Authors
Citation
Xg. Liang, THE BOUNDARY INDUCED ERROR ON THE MEASUREMENT OF THERMAL-CONDUCTIVITYBY TRANSIENT HOT-WIRE METHOD, Measurement science & technology, 6(5), 1995, pp. 467-471
Citations number
9
Categorie Soggetti
Instument & Instrumentation",Engineering
ISSN journal
09570233
Volume
6
Issue
5
Year of publication
1995
Pages
467 - 471
Database
ISI
SICI code
0957-0233(1995)6:5<467:TBIEOT>2.0.ZU;2-Y
Abstract
An analysis of the dimension influence on the mesurement accuracy of t hermal conductivities by transient hot wire/probe method is made. The radial and axial effects are discussed separately under the suppositio n of constant boundary temperatures. The results show that the radial and axial lengths of the measured sample, the radius ratio and heat ca pacity ratio of sample to hot wire/probe, as well as the length to dia meter ratio of the hot wire/probe are important parameters in error an alysis. Former analysis underestimated the boundary effects. New crite ria for error analysis are established.