Lp. Loginova et Vd. Bazilyanskaya, IODIDE BUFFER SYSTEMS FOR THE CALIBRATION OF IODIDE-SELECTIVE ELECTRODES, Analytica chimica acta, 307(1), 1995, pp. 37-41
A new approach to the creation of anion buffer systems is proposed, wh
ich extends the linear range. The new iodide buffer systems, containin
g (1) HgI2 precipitate, CaEDTA(2-), Ca2+, HgEDTA(2-), and (2) HgI2 pre
cipitate, chloride complexes of Hg2+ and excess of Cl- ions, are propo
sed for the calibration of iodide-selective electrodes at pI > 5. It i
s shown with the help of these buffers, that a solid-state Orion 94-53
iodide-selective electrode keeps linear Nemstian response up to pI =
7.5. Iodide buffer systems are used for the estimation of the reliabil
ity of literature data on selectivity coefficients of iodide-selective
electrodes with respect to Cl- ions.