RUGGEDNESS TESTING OF A SIZE-EXCLUSION CHROMATOGRAPHIC ASSAY FOR LOW-MOLECULAR-MASS POLYMERS

Citation
Y. Vanderheyden et al., RUGGEDNESS TESTING OF A SIZE-EXCLUSION CHROMATOGRAPHIC ASSAY FOR LOW-MOLECULAR-MASS POLYMERS, Journal of chromatography, 756(1-2), 1996, pp. 89-106
Citations number
18
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
Journal title
Volume
756
Issue
1-2
Year of publication
1996
Pages
89 - 106
Database
ISI
SICI code
Abstract
A ruggedness test of a size-exclusion chromatographic assay for a low- molecular-mass polymer mixture was performed by applying a fractional factorial design. Eight factors selected from the method procedure wer e examined at three levels by reflecting the design. The effects of th e factors were calculated and interpreted, statistically as well as gr aphically. The statistical interpretation method based on the use of t wo-factor interaction effects to estimate experimental error was found to be effective to indicate significant effects in a ruggedness test. The factors column manufacturer and detector type turned out to be th e least robust. Variations in the other factors, within the levels exa mined, do not lead to chromatographically relevant changes.