B. Kerstingsommerhoff et al., MRI OF KNEE-JOINT - FIRST RESULTS OF A CO MPARISON OF A 0.2 T DEDICATED SYSTEM WITH A 1.5 T HIGH-FIELD STRENGTH MAGNET, RoFo. Fortschritte auf dem Gebiete der Rontgenstrahlen und der neuenbildgebenden Verfahren, 162(5), 1995, pp. 390-395
Purpose: Diagnostic accuracy and image quality of a specialised system
for MR examinations of peripheral joints were evaluated. Material and
Methods: 20 patients with acute or chronic injuries of the knee were
examined using a low-field MR system (0.2 T). For comparison, all pati
ents were also studied with a 1.5 T high field strength magnet and all
diagnoses were correlated with arthroscopic findings. Results: We fou
nd compatible diagnostic accuracies (cruciate ligaments 90 %, menisci
75-90 %) and good image quality ratings for the low field system (''go
od'' and ''excellent'' image quality in 83 % of cases). Conclusion: Th
e low-field MR-system offers low-cost MR examinations of peripheral jo
ints with good image quality and reliable diagnostic information.