Troponin T (TnT) is recently being considered to be an important diagn
ostic marker of myocardial damage in adults, but this marker has not y
et been used in neonates. The present study was designed to determine
the normal level of cardiac TnT in the cord blood of healthy term neon
ates. Cardiac troponin T concentration in cord blood was measured in 1
5 healthy term neonates using commercial kit (Enzymun-Test System, Boe
hringer, Mannheim). TnT serum concentration was 0.05+/-0.04 mu g/l in
10 of 15 babies whereas in the remaining 5 haemolysed samples its conc
entration was elevated (mean 0.19+/-0.07 mu g/l). It is important to c
onsider that incidental haemolysis of blood samples can mimic patholog
ical elevation of TnT by interfering with the assay.