Rf. Steinert et al., LONG-TERM CLINICAL-RESULTS OF AMO PHACOFLEX MODEL SI-18 INTRAOCULAR-LENS IMPLANTATION, Journal of cataract and refractive surgery, 21(3), 1995, pp. 331-338
The AMO (R) PhacoFlex (R) Model SI-18 was the first commercially avail
able three-piece silicone intraocular lens (IOL) for use in small inci
sion cataract surgery. Allergan Medical Optics' silicone IOLs have bee
n implanted in more than 750,000 patients worldwide over the past eigh
t years. This report of the FDA clinical investigation of the AMO Phac
oFlex model SI-18 IOL summarizes one year follow-up data from the init
ial premarket approval clinical trial on 500 core patients implanted w
ith the SI-18 lens. One-year follow-up data from 5,860 patients in the
modified core group and three-year follow-up data on the 500 patient
core/modified core group are also presented. At three years postoperat
ively, 91.3% of best case core/modified core patients achieved 20/40 o
r better corrected visual acuity. Overall incidence of persistent comp
lications at three years was 3.2%.