ATMOSPHERIC AEROSOL SAMPLING BY AN ANNULAR DENUDER SYSTEM AND A HIGH-VOLUME PM(10) SAMPLER

Authors
Citation
Cj. Tsai et Hy. Huang, ATMOSPHERIC AEROSOL SAMPLING BY AN ANNULAR DENUDER SYSTEM AND A HIGH-VOLUME PM(10) SAMPLER, Environment international, 21(3), 1995, pp. 283-291
Citations number
NO
Categorie Soggetti
Environmental Sciences
Journal title
ISSN journal
01604120
Volume
21
Issue
3
Year of publication
1995
Pages
283 - 291
Database
ISI
SICI code
0160-4120(1995)21:3<283:AASBAA>2.0.ZU;2-R
Abstract
An Annular Denuder System (ADS) was used to study sampling errors of t otal PM(10) and soluble ionic species concentrations determined by the high volume PM(10) (HVPM(10)) sampler. Experimental results show that in the HVPM(10) sampler, more alkaline glass fiber filters collect mo re sulfate and nitrate but less ammonium ion than quartz filters. Howe ver, total PM(10) concentrations do not seem to depend on the types of filters. Artifact sulfate concentrations in glass fiber filters of th e HVPM(10) sampler are found to exceed those predicted by the Coutant' s model. It is suggested that artifact sulfate and nitrate not only oc cur within filters but also in particle deposits. Chloride ion concent rations in the filters of the HVPM(10) sampler are underestimated sign ificantly due to evaporative loss of ammonium chloride.