COUPLING OF THE RECOIL MASS-SPECTROMETER CAMEL TO THE GAMMA-RAY SPECTROMETER GASP

Citation
P. Spolaore et al., COUPLING OF THE RECOIL MASS-SPECTROMETER CAMEL TO THE GAMMA-RAY SPECTROMETER GASP, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 359(3), 1995, pp. 500-505
Citations number
10
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
359
Issue
3
Year of publication
1995
Pages
500 - 505
Database
ISI
SICI code
0168-9002(1995)359:3<500:COTRMC>2.0.ZU;2-R
Abstract
A project has been realized to link the CAMEL recoil mass spectrometer to the GASP gamma-spectrometer in order to perform high resolution an d efficiency gamma-recoil coincidence measurements. To preserve high f lexibility and autonomy in the operation of the two complex apparatus a rough factor two of reduction in the overall heavy ion transmission was accepted in designing the optics of the particle transport from th e GASP center to the CAMEL focal plane. The coupled configuration has been tested with the fusion reaction Ni-58 (E = 212 MeV) + Ni-64, Obta ining a mass resolution of 1/300 and efficiency between similar to 11% and similar to 15% for different evaporation products.