ALPHA-RADIATION DETECTION AND ANALYSIS BY NUCLEAR TRACK DETECTOR

Citation
G. Espinosa et al., ALPHA-RADIATION DETECTION AND ANALYSIS BY NUCLEAR TRACK DETECTOR, Radiation protection dosimetry, 59(3), 1995, pp. 227-229
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
ISSN journal
01448420
Volume
59
Issue
3
Year of publication
1995
Pages
227 - 229
Database
ISI
SICI code
0144-8420(1995)59:3<227:ADAABN>2.0.ZU;2-8
Abstract
A gross analysis is presented of alpha radiation from Am-241, Pu-239, Th-230, Cm-244, U-238 and Np-237 sources, using nuclear track detector (NTD) methods. The response of the NTD was measured at low and high a lpha emission rates, fixed exposure times, and fixed quantity of total alpha particles for each of the different isotopes over a range of al pha energies. The distributions and statistics of the track densities were calculated. LANTRAK polycarbonate was selected as the NTD detecto r material. A one-step chemical etch in 6.25N NaOH at 60-degrees-C for 16 h was used. Twenty-four fields (0.91 mm2) one each detector were r ead using an automatic counting system. The results show the linearity and efficiency of the material for each alpha energy, and the track d iameter distribution relative to energy.