A rapid simple and robust reversed-phase HPLC method was developed for
rapid screening in bioavailability studies or comparative bioequivale
nce studies. The method is specific for vancomycin as no interference
from acetylsalicylic acid, paracetamol and caffeine was observed. The
mean intra-day precision was from 11.7% (low concentration) to 0.3% (h
igh concentration) and the within-day precision from 15.0 to 0.3%, det
ermined on spiked samples. The accuracy of the method was 106.4-99.8%
(intra-day) and 103.5-100.2% (inter-day).