Jc. Gao et Zy. Wang, THE EXTENSION OF THE RANGE OF NIM ALANINE ESR DOSIMETRIC SYSTEM TO THERAPY LEVEL/, Applied radiation and isotopes, 47(11-12), 1996, pp. 1193-1196
Citations number
4
Categorie Soggetti
Nuclear Sciences & Tecnology","Radiology,Nuclear Medicine & Medical Imaging","Chemistry Inorganic & Nuclear
The NIM alanine/ESR dosimetric system, which was designed for industri
al radiation processing, has been improved to be suited for applicatio
ns in therapy dose range. Two different procedures of dose intercompar
isons between IAEA and NIM were carried out with the improved system i
n the range of 2.5 to 100 Gy. In the first procedure, a set: of NIM al
anine dosimeters were irradiated at IAEA dosimetric laboratory and par
t of dosimeters marked ''for evaluated'' were evaluated using the rest
of those with ''known dose'' given by IAEA. Most of evaluated doses a
greed with IAEA doses within 1%. In the second procedure, all above do
simeters were evaluated on the base of NIM dosimetry. The results indi
cated that the doses determined by NIM agreed with those given by IAEA
within 3% on the average. Copyright (C) 1996 Elsevier Science Ltd