The suitability of the ESR alanine and sugar detectors for criticality
accident dosimetry was experimentally investigated during an intercom
parison of dosimetry techniques. Tests were performed irradiating dete
ctors both free-in-air and on-phantom during controlled criticality ex
cursions at the SILENE reactor in Valduc, France. Several grays of abs
orbed dose Were imparted in neutron gamma-ray fields of various relati
ve intensities and spectral distributions. Analysed results confirmed
the potential of these systems which can immediately provide an acute
dose assessment with an average underestimate of 30% in the various fi
elds. This performance allows for the Screening of severely exposed in
dividuals and meets the IAEA recommendations on the early estimate Of
accident absorbed doses. Copyright (C) 1996 Elsevier Science Ltd