Pg. Bjornstad, WIRE TRAPPING IN RASHKINDS PATENT DUCTUS-ARTERIOSUS OCCLUDER - ITS ETIOLOGY AND HOW IT MAY BE AVOIDED, Cardiovascular and interventional radiology, 18(3), 1995, pp. 203-204
Release of the Rashkind ductus umbrella may be complicated in that the
central wire of the delivery catheter may become trapped within the f
oam of the umbrella. This problem occurs only if the central retaining
wire is advanced when trying to disengage the catheter and umbrella.
We propose a simple solution to the problem: Retract the sleeve instea
d of advancing the retaining wire.