WIRE TRAPPING IN RASHKINDS PATENT DUCTUS-ARTERIOSUS OCCLUDER - ITS ETIOLOGY AND HOW IT MAY BE AVOIDED

Authors
Citation
Pg. Bjornstad, WIRE TRAPPING IN RASHKINDS PATENT DUCTUS-ARTERIOSUS OCCLUDER - ITS ETIOLOGY AND HOW IT MAY BE AVOIDED, Cardiovascular and interventional radiology, 18(3), 1995, pp. 203-204
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
01741551
Volume
18
Issue
3
Year of publication
1995
Pages
203 - 204
Database
ISI
SICI code
0174-1551(1995)18:3<203:WTIRPD>2.0.ZU;2-#
Abstract
Release of the Rashkind ductus umbrella may be complicated in that the central wire of the delivery catheter may become trapped within the f oam of the umbrella. This problem occurs only if the central retaining wire is advanced when trying to disengage the catheter and umbrella. We propose a simple solution to the problem: Retract the sleeve instea d of advancing the retaining wire.