K. Hynynen et al., CYLINDRICAL ULTRASONIC TRANSDUCERS FOR CARDIAC CATHETER ABLATION, IEEE transactions on biomedical engineering, 44(2), 1997, pp. 144-151
This study was designed to evaluate the feasibility of using cylindric
al ultrasound transducers mounted on a catheter for the ablation of ca
rdiac tissues, In addition, the effects of ultrasound frequency and po
wer was evaluated both using computer simulations and in vitro experim
ents, Frequencies of 4.5, 6, and 10 MHz were selected based on the sim
ulation studies and manufacturing feasibility, These transducers were
mounted on the tip of 7-French catheters and applied in vitro to fresh
ventricular canine endocardium, submerged in flowing degassed saline
at 37 degrees C, When the power was regulated to maintain transducer i
nterface temperature at 90-100 degrees C, the 10-, 6-, and 4.5-MHz tra
nsducers generated a lesion depth of 5.9 +/- 0.2 mm, 4.6 +/- 1.0 mm, a
nd 5.3 +/- 0.9 mm, respectively. The 10-MHz transducer was chosen for
the in vivo tests since the maximum lesion depth was achieved with the
lowest power. Two dogs mere anesthetized and sonications were perform
ed in both the left and right ventricles, The 10-MHz cylindrical trans
ducers caused an average lesion depth of 6.4 +/- 2.5 mm, In conclusion
, the results show that cylindrical ultrasound transducers can be used
for cardiac tissue ablation and that they may be able to produce deep
er tissue necrosis than other methods currently in use.