F. Folkmann et Hr. Nagel, MEASUREMENTS OF X-RAYS FROM NE RECOIL IONS WITH A POSITION-SENSITIVE DETECTOR, Nuclear instruments & methods in physics research. Section B, Beam interactions with materials and atoms, 98(1-4), 1995, pp. 589-592
Measurements of soft X-rays with wavelength 35-47 nm from Ne recoil io
ns, excited by 24 MeV Cl ions, are reported obtained with a 2.2 m graz
ing incidence spectrometer. A position sensitive channelplate detector
with a two-dimensional wedge and strip readout is used instead of a c
hanneltron, scanning along the Rowland circle. The perfomance of this
position sensitive detector is optimized, wavelength calibration proce
dures are worked out and its efficiency is compared with the scanning
channeltron. The detection yield and resolution is measured as a funct
ion of the angle of the radiation to the detector normal. An increase
by a factor of 12 in the detection efficiency over a spectral region i
s found for the position sensitive detector compared with a scanning c
hanneltron detector.