Fa. Bencsath et al., OPTIMIZATION OF THE ANALYTICAL PERFORMANCE OF THE MAGNETIC-SECTOR MASS-SPECTROMETER FOR THE IDENTIFICATION OF RESIDUAL CHLORAMPHENICOL IN SHRIMP, Biological mass spectrometry, 23(11), 1994, pp. 665-674
Chemical noise limits mass spectrometric detection of chloramphenicol
(CAP) with electron capture ionization at low resolution, and makes CA
P identification at concentrations of 5 parts per billion (ppb) diffic
ult. Increasing the resolution from 1000 to 3500, however, was suffici
ent to separate the analyte signals from the noise signals, and result
ed in a 100 times higher analytical sensitivity. The introduction of s
weep gas in the ion source decreased the scattering of the quantitativ
e results on average by a factor of 7, and thereby improved the precis
ion of the analyses to an acceptable level (CV < 10%). Under such cond
itions, CAP residues of 1.5 and 2.1 ppb in shrimp as determined by ele
ctron capture gas chromatography/mass spectrometry can readily be iden
tified by monitoring four diagnostic ions.