DETERMINATION OF IRON AT NG ML LEVEL BY SOLID-PHASE SPECTROPHOTOMETRYAFTER PRECONCENTRATION ON CATION-EXCHANGE FILTERS/

Citation
Fc. Garcia et al., DETERMINATION OF IRON AT NG ML LEVEL BY SOLID-PHASE SPECTROPHOTOMETRYAFTER PRECONCENTRATION ON CATION-EXCHANGE FILTERS/, Talanta, 42(5), 1995, pp. 711-715
Citations number
17
Journal title
Talanta
ISSN journal
00399140 → ACNP
Volume
42
Issue
5
Year of publication
1995
Pages
711 - 715
Database
ISI
SICI code
0039-9140(1995)42:5<711:DOIANM>2.0.ZU;2-O
Abstract
A cationic exchanger paper is used to retain analytes in solution and, after drying, to analyze directly by measuring the UV-Vis absorbance of the paper. The method was applied to determination of iron using it s known 1,10-phenanthroline complex. Using 100 ml of sample the applic able concentration range was between 1.0 and 10.0 ng/ml with a detecti on limit of 0.2 ng/ml and a RSD around 2%. The method was applied to d etermination of iron total in snow, human serum and wine.