Fc. Garcia et al., DETERMINATION OF IRON AT NG ML LEVEL BY SOLID-PHASE SPECTROPHOTOMETRYAFTER PRECONCENTRATION ON CATION-EXCHANGE FILTERS/, Talanta, 42(5), 1995, pp. 711-715
A cationic exchanger paper is used to retain analytes in solution and,
after drying, to analyze directly by measuring the UV-Vis absorbance
of the paper. The method was applied to determination of iron using it
s known 1,10-phenanthroline complex. Using 100 ml of sample the applic
able concentration range was between 1.0 and 10.0 ng/ml with a detecti
on limit of 0.2 ng/ml and a RSD around 2%. The method was applied to d
etermination of iron total in snow, human serum and wine.