Rm. Mcmurphy et al., MODIFICATION OF A NONREBREATHING CIRCUIT ADAPTER TO PREVENT BAROTRAUMA IN ANESTHETIZED PATIENTS, Veterinary surgery, 24(4), 1995, pp. 352-355
Barotrauma, pneumothorax, and pneumomediastinum occurred in two anesth
etized cats in which the waste gas outlet of a nonrebreathing circuit
was occluded. To prevent any similar cases of barotrauma, we have modi
fied our nonrebreathing circuit adapters by inserting a 15 cm H2O PEEP
valve into the gas pathway of the nonrebreathing circuit adapter. Thi
s PEEP valve prevents the circuit and airway pressures from exceeding
15 cm H2O if the pop-off valve of the nonrebreathing circuit adapter i
s inadvertently left closed. (C)Copyright 1995 by The American College
of Veterinary Surgeons