MODIFICATION OF A NONREBREATHING CIRCUIT ADAPTER TO PREVENT BAROTRAUMA IN ANESTHETIZED PATIENTS

Citation
Rm. Mcmurphy et al., MODIFICATION OF A NONREBREATHING CIRCUIT ADAPTER TO PREVENT BAROTRAUMA IN ANESTHETIZED PATIENTS, Veterinary surgery, 24(4), 1995, pp. 352-355
Citations number
7
Categorie Soggetti
Veterinary Sciences
Journal title
ISSN journal
01613499
Volume
24
Issue
4
Year of publication
1995
Pages
352 - 355
Database
ISI
SICI code
0161-3499(1995)24:4<352:MOANCA>2.0.ZU;2-#
Abstract
Barotrauma, pneumothorax, and pneumomediastinum occurred in two anesth etized cats in which the waste gas outlet of a nonrebreathing circuit was occluded. To prevent any similar cases of barotrauma, we have modi fied our nonrebreathing circuit adapters by inserting a 15 cm H2O PEEP valve into the gas pathway of the nonrebreathing circuit adapter. Thi s PEEP valve prevents the circuit and airway pressures from exceeding 15 cm H2O if the pop-off valve of the nonrebreathing circuit adapter i s inadvertently left closed. (C)Copyright 1995 by The American College of Veterinary Surgeons