A quartz crystal microbalance (QCM) bioaffinity sensor for biotin, bas
ed on the bioaffinity complex between avidin and immobilized desthiobi
otin, was established. BSA-desthiobiotin conjugate was adsorbed direct
ly onto the gold layer surface of a quartz crystal. The crystal was pl
aced in a QCM flow cell. Avidin formed a metastable complex with the B
SA-desthiobiotin conjugate on the crystal surface. When a sample solut
ion containing biotin was injected, the avidin was released from the s
urface to form a more stable complex with biotin in solution. The biot
in concentration corresponded to the weight change on the crystal surf
ace that occurred when avidin desorbed. Measurements were done in the
stop-flow mode with a measurement range between 0.05 and 100 mu g/mL a
nd in the continuous-flow mode with a measurement range between 0.2 an
d 10 000 mu g/mL. The sensor was regenerated without any loss of activ
ity by readsorbing avidin.