INCONSISTENT RESPONSE TO RAPID ATRIAL RHYTHMS IN A DDD PACEMAKER WITHTHE FALLBACK FEATURE

Authors
Citation
Nj. Johnson, INCONSISTENT RESPONSE TO RAPID ATRIAL RHYTHMS IN A DDD PACEMAKER WITHTHE FALLBACK FEATURE, PACE, 18(7), 1995, pp. 1458-1462
Citations number
3
Categorie Soggetti
Cardiac & Cardiovascular System","Engineering, Biomedical
ISSN journal
01478389
Volume
18
Issue
7
Year of publication
1995
Pages
1458 - 1462
Database
ISI
SICI code
0147-8389(1995)18:7<1458:IRTRAR>2.0.ZU;2-F
Abstract
Dual chamber pacing may be problematic in patients with paroxysmal atr ial arrhythmias. The fallback feature has been incorporated into-some devices to avoid tracking of rapid atrial rhythms. A case is reported in which the initiation of fallback was inconsistent during paroxysms of atrial flutter. Evaluation of the ventricular output of-an isolated device in response to various applied atrial rates revealed that fall back occurred consistently only when 2xTARP (ms) less than or equal to URL (ms).