Fl. Thaete et al., REPRODUCIBILITY OF COMPUTED-TOMOGRAPHY MEASUREMENT OF VISCERAL ADIPOSE-TISSUE AREA, International journal of obesity, 19(7), 1995, pp. 464-467
OBJECTIVE: The purpose of the current study is to examine the reproduc
ibility of CT measurement of the cross-sectional area of abdominal adi
pose tissue. DESIGN: Sixteen healthy pre-menopausal women, ranging fro
m lean to obese, underwent duplicate cross-sectional CT scans of the a
bdomen at the level of the 4th lumbar vertebral disc space. Subjects s
tood after the initial scan and were then re-positioned prior to repea
t CT scanning. RESULTS: There was a high degree of reproducibility in
measurement of overall cross-sectional adipose tissue (AT) and in the
partitioning of this into subcutaneous and visceral AT. The excellent
reproducibility is demonstrated by a high correlation between duplicat
e measurements (r = 0.99, P < 0.01), and by small precision errors; 1.
2% of the mean value for total AT cross-sectional area, 1.9% for subcu
taneous AT area, and 3.9% for visceral AT area. CONCLUSION: Measuremen
t of abdominal AT area by cross-sectional CT scanning is not only an a
ccurate determination of this adipose depot but can also be repeated w
ith a high degree of precision.