FREQUENCY-MODULATED SIMULTANEOUS MULTIELEMENT ATOMIC-ABSORPTION SPECTROMETRY USING ELECTROTHERMAL ATOMIZER AND DEUTERIUM BACKGROUND CORRECTION

Citation
H. Edel et al., FREQUENCY-MODULATED SIMULTANEOUS MULTIELEMENT ATOMIC-ABSORPTION SPECTROMETRY USING ELECTROTHERMAL ATOMIZER AND DEUTERIUM BACKGROUND CORRECTION, Fresenius' journal of analytical chemistry, 351(6), 1995, pp. 479-483
Citations number
12
Categorie Soggetti
Chemistry Analytical
ISSN journal
09370633
Volume
351
Issue
6
Year of publication
1995
Pages
479 - 483
Database
ISI
SICI code
0937-0633(1995)351:6<479:FSMAS>2.0.ZU;2-V
Abstract
Characteristic data of the frequency-modulated simultaneous multieleme nt atomic absorption spectrometry (FREMSAAS) using electrothermal atom izer with deuterium background correction, have been determined. The d ata obtained have been processed by using several statistical tests re commended for quality control purposes. The instrumentation has been p resented as well as procedures of separating elements into measure gro ups and fixing of variable conditions. Detection limits, characteristi c masses and working ranges have been given for the eleven elements ex amined. The data have been in good agreement to results obtained with conventional one-channel AAS instruments. The eleven elements have bee n simultaneously determined in a standard reference material (SRM) and all results are compatible with a 95% certainty with the certified va lues. FREMSAAS has been applied to a real sample.