The determination of iodine, bromine, sulphur and phosphorus using dif
ferent modes of sampling in ICP-AES is described. Detection limits in
the range of some 10 mu gL(-1) have been obtained using directly aqueo
us solutions with pneumatic nebulization. The influence of several ani
ons and metals has been investigated. The addition of methanol and eth
anol have improved sensitivities but decreased signal-to-background ra
tios and detection limits due to extended blank values and increased s
tandard deviations. Using ultrasonic nebulization the and the detectio
n limits could be enhanced for these elements in aqueous solutions, bu
t a significant decrease in sensitivity has been observed using additi
ves as methanol or ethanol. By coupling of pneumatic nebulization and
gas generation of iodine by addition of oxidizing agents, the detectio
n limits of iodine could be improved to 6 ngL(-1) at the 178.28 nm lin
e. Electrothermal vaporization has successfully been applied for micro
sampling of nonmetals and ICP-AES determination. Using aqueous solutio
ns without any modifier the following absolute detection limits at the
S (180.73 nm), P (178.27 nm), I (178.28 nm) and Br (163.34 nm) lines
have been obtained: 0.25 ng of S, 0.44 ng of P, 0.36 ng of I and 43 ng
of Br.