The convex bileaflet valve replaces the flat biflap inflow valve desig
ned by Long Sheng Yu and the tricusp semilunair outflow valve. One rea
son is easier manufacturing. Convex bileaflet valves are developed for
the 11, 20, 40, 70 and 140cc ventricles. Testing included curves (Car
diac Output versus Venous Pressure, Cardiac Output versus Heart rate),
flow visualization studies, paint and bloodbag studies. The curves an
d flow visualization were done by connecting ventricles to one of our
standard mock circulations. Paint and bloodbag studies were done by co
nnecting the hearts to a bloodbag, but the bag was filled with water f
or the paint studies. The curves show high cardiac output, even with p
umping at high heart rates (150 BPM+). The flow visualization shows a
good stream through the sinus Valsalvae. No stagnating flow is visible
. The bloodbag studies which provoke thrombosis show it on the edges o
f the heart valves, and little in the groove between the valve and the
sinus Valsalvae. Heparninzation prevents the thrombosis. Results of o
ur tests were good. The convex bileaflet valve seems to have good futu
re.