AMPEROMETRIC DETERMINATION OF BILIRUBIN WITH FLOW-INJECTION ANALYSIS SYSTEM

Citation
Zu. Bae et al., AMPEROMETRIC DETERMINATION OF BILIRUBIN WITH FLOW-INJECTION ANALYSIS SYSTEM, Analytical letters, 28(10), 1995, pp. 1775-1783
Citations number
21
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
00032719
Volume
28
Issue
10
Year of publication
1995
Pages
1775 - 1783
Database
ISI
SICI code
0003-2719(1995)28:10<1775:ADOBWF>2.0.ZU;2-2
Abstract
The amperometric method using the flow injection system has been devel oped for the determination of bilirubin. Bilirubin is oxidized to bili verdin at the glassy carbon working electrode. The optimum conditions were investigated. Linear calibration curve was obtained between 1.0x1 0(-6) and 1.0x10(-3)M, with a sampling rate of 20 samples h(-1) and a relative standard deviation of 3.2%. The limit of detection is 4.0x10( -7) M. Interference of hemoglobin was not observed.