ABSOLUTE TC-99(M)-DMSA RENAL UPTAKE IN CHILDREN - OPTIMUM TIME TO SCAN

Citation
Ma. Flower et al., ABSOLUTE TC-99(M)-DMSA RENAL UPTAKE IN CHILDREN - OPTIMUM TIME TO SCAN, Nuclear medicine communications, 16(7), 1995, pp. 572-574
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
01433636
Volume
16
Issue
7
Year of publication
1995
Pages
572 - 574
Database
ISI
SICI code
0143-3636(1995)16:7<572:ATRUIC>2.0.ZU;2-P
Abstract
Absolute renal uptake was measured at 2, 4 and 6 h in 27 patients in o rder to determine a more convenient time for uptake compared with the 6-h measurement proposed by other authors. Measurements made at 2 and 4 h would need to be increased by 20 and 6%, respectively, to convert to the value at 6 h. Measurements at 4 h are a reasonable compromise t o achieve a high-count/low-background image in a reasonable time and t o obtain a good estimate of the predicted 6-h uptake at a single scann ing session. The percent renal uptake at 6 h, U(6), can be derived fro m the percent uptake measured at t hours after injection, U(t), using the following formula: U(6) = U(t) x CF(t), where CF(t) = 1.0 + 0.03 ( 6 - t) for 4 less than or equal to t less than or equal to 6.