Ah. Maitz et al., QUALITY ASSURANCE FOR GAMMA-KNIFE STEREOTAXIC RADIOSURGERY, International journal of radiation oncology, biology, physics, 32(5), 1995, pp. 1465-1471
Citations number
9
Categorie Soggetti
Oncology,"Radiology,Nuclear Medicine & Medical Imaging
Purpose: This quality assurance program is designed for stereotactic r
adiosurgical units, gamma knife, to check and maintain the unit to pre
clude accidents and comply with current regulations. Materials and Met
hods: Over 58 stereotactic radiosurgical units using 201 focused (CO)-
C-60 beams have been installed in the last 7 years and are in use at h
ospitals throughout the world, with at least 11 additional units being
prepared to come on-line in the next year, This system has been in us
e at the University of Pittsburgh Medical Center (UPMC) for 7 years, A
comprehensive quality assurance program has been developed, It includ
es the physics and dosimetry parameters and safety checks required by
regulatory agencies, The program, based on over 7 years of experience
in measurements, and used during the treatment of over 1500 patients,
is separated into three aspects, namely physics, dosimetry, and safety
, The UPMC program hopefully will indicate out-of-tolerance problems,
Some quality assurance items are checked on a daily basis prior to pat
ient treatment, while other aspects are checked on a weekly, monthly,
and/or annual basis, A complete list of items with their respective ti
me tables and tolerances is provided. Results: Although experience sho
ws very small margins of error, larger values were chosen to account f
or variations in equipment and techniques, Conclusions: Items included
in this quality assurance program should indicate and/or preclude pro
blems encountered in the use of this unit.