Patients with cardiac pacemakers are currently restricted from nuclear
magnetic resonance imaging (MRI). The aim of the study was to analyze
the influence of MRI on new generation pacemakers. Tests were perform
ed using a phantom model with seven dual chamber and two single chambe
r systems in a 0.5 Tesla MRI scanner. Monitoring by telemetry and osci
llography were used during the standard clinical scan sequences as wel
l as a pacemaker inquiry after each sequence. Spin echo, gradient echo
, and fast field echo sequences were performed with the following stim
ulation modes: VVI, VVIR, VOO, DDD, DDDR, and DOO. On entering the sta
tic magnetic field, the reed switch was activated followed by asynchro
nous stimulation. The subsequent scan showed no influence on the stimu
lation function nor on the pacemaker program. Event counter function r
emained intact. Pacemakers with automatic mode switching to demand pac
ing or programmed inactivation of the reed switch were triggered in th
e dual chamber mode and were inhibited in the one chamber mode during
the scan. Alterations of pacemaker program or rapid pacing were not ob
served. MRI scan could induce voltage as high as intracardiac signals,
but the stimulation threshold of the heart was not reached. Thus, pac
emakers should be programmed in the asynchronous mode during scan to a
void inhibition and trigger mechanism.