Flow injection analysis (FIA) systems incorporating potentiometric sen
sors for harmine and harmaline hallucinogens are described. The sensor
s are based on the use of harmine and harmaline tetraphenylborate or r
eineckate as electroactive compounds dispersed in plasticized PVC matr
ix. Laboratory-build flow-through sandwich cell equipped with silver d
isk coated with the sensing material is used as a detector in a low di
spersion FIA system. The detector exhibits fast and linear response ov
er the concentration range 10(-2)-10(-6) M harmine or harmaline with a
detection limit of 2 mu g/mL (for 100 mu L samples). Operational pH r
anges from 3 to 8, sampling rate is 80 per hour and the precision is t
ypically +/- 0.6%. Recovery and reproducibility data agree fairly well
with those obtained by gas-liquid chromatography.