PERFORMANCE OF A GAS-CHROMATOGRAPH ATOMIC-EMISSION DETECTOR SYSTEM INA MOBILE ANALYTICAL LABORATORY

Citation
Gj. Hudak et al., PERFORMANCE OF A GAS-CHROMATOGRAPH ATOMIC-EMISSION DETECTOR SYSTEM INA MOBILE ANALYTICAL LABORATORY, Journal of hazardous materials, 43(1-2), 1995, pp. 155-167
Citations number
3
Categorie Soggetti
Engineering, Civil","Environmental Sciences","Engineering, Environmental
ISSN journal
03043894
Volume
43
Issue
1-2
Year of publication
1995
Pages
155 - 167
Database
ISI
SICI code
0304-3894(1995)43:1-2<155:POAGAD>2.0.ZU;2-0
Abstract
The effect of ambient environmental conditions on a gas chromatography -atomic emission detection (GC-AED) system was characterized. The envi ronmental stimuli was shock and vibration due to operating the hardwar e in a mobile analytical laboratory. The effect of large temperature c hange was also examined. System performance using a reference sample a nd operational recommendations are given.