A BIOSENSOR AS WARNING DEVICE FOR THE DETECTION OF CYANIDE, CHLOROPHENOLS, ATRAZINE AND CARBAMATE PESTICIDES

Citation
Jl. Besombes et al., A BIOSENSOR AS WARNING DEVICE FOR THE DETECTION OF CYANIDE, CHLOROPHENOLS, ATRAZINE AND CARBAMATE PESTICIDES, Analytica chimica acta, 311(3), 1995, pp. 255-263
Citations number
19
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
00032670
Volume
311
Issue
3
Year of publication
1995
Pages
255 - 263
Database
ISI
SICI code
0003-2670(1995)311:3<255:ABAWDF>2.0.ZU;2-D
Abstract
The determination of cyanide, chlorophenols, atrazine, dithiocarbamate and carbamate pesticides is described, utilizing an amperometric bios ensor constructed by the electropolymerization of a pyrrole amphiphili c monomer-tyrosinase coating. Measurements were carried out with catec hol, dopamine, L-DOPA or epinephrine as an enzyme substrate; the enzym atically generated quinoid products being electroreduced at -200 mV vs . SCE. The detection of these water pollutants was performed via their inhibiting action on the tyrosinase electrode. The characterization o f the inhibition processes (competitive/non-competitive) and their rev ersibility were examined. The detection limits are 0.4, 2, 2, 4 and 0. 02 mu M for 3,4-dichlorophenol, chloroisopropylphenylcarbamate, 3-chlo roaniline, atrazine and cyanide, respectively.