Ms. Huq et al., AN EVALUATION OF THE RECOMMENDATIONS OF THE TG-25 PROTOCOL FOR DETERMINATION OF DEPTH-DOSE CURVES FOR ELECTRON-BEAMS USING IONIZATION CHAMBERS, Medical physics, 22(8), 1995, pp. 1333-1337
AAPM Radiation Therapy Committee Task Group 25 has recently outlined a
protocol for the determination of relative dose curves for electron b
eams [Med. Phys. 18, 73-109 (1991)]. We have performed an evaluation o
f this protocol by comparing the central axis depth dose curves determ
ined from measurements using two different ionization chambers and thr
ee different phantom materials. Measurements were made with a Farmer-t
ype PTW and Capintec ionization chamber in solid water, PMMA, and clea
r polystyrene phantoms irradiated by 6- and 15-MeV electron beams. Cen
tral axis depth dose curves were generated from the measured depth-ion
ization data using the new protocol. For both the chambers and energie
s investigated in this study, excellent agreement was observed among a
ll the depth doses in water obtained from measurements in all of the t
hree phantoms studied.