A COMPACT PARALLEL-PLATE DETECTOR FOR HEAVY-ION REACTION STUDIES

Citation
S. Beghini et al., A COMPACT PARALLEL-PLATE DETECTOR FOR HEAVY-ION REACTION STUDIES, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 362(2-3), 1995, pp. 526-531
Citations number
13
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
362
Issue
2-3
Year of publication
1995
Pages
526 - 531
Database
ISI
SICI code
0168-9002(1995)362:2-3<526:ACPDFH>2.0.ZU;2-M
Abstract
A two dimensional position sensitive parallel plate detector, with an active area of 10 X 8 cm(2), was developed for heavy ion reaction stud ies. Position signals are extracted by using a non-standard compact de lay-line, which allows saving dead spaces. The timing (less than or eq ual to 280 ps) and position (1 mm in X, Y) resolutions of the detector make it a useful device for kinematic coincidence studies. Moreover, due to its mechanical construction, it is well suited for the implemen tation of a 4 pi structure to be built e.g. in the inner space of a Ge -array.