SYNCHROTRON EXPERIMENTAL RESULTS ON AN UNRIBBED SAGITTALLY FOCUSING CRYSTAL MONOCHROMATOR

Citation
Jp. Quintana et al., SYNCHROTRON EXPERIMENTAL RESULTS ON AN UNRIBBED SAGITTALLY FOCUSING CRYSTAL MONOCHROMATOR, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 362(2-3), 1995, pp. 592-594
Citations number
8
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
362
Issue
2-3
Year of publication
1995
Pages
592 - 594
Database
ISI
SICI code
0168-9002(1995)362:2-3<592:SEROAU>2.0.ZU;2-N
Abstract
Results are reported on the performance of a novel sagittally focussin g crystal monochromator design that minimizes the anticlastic bending without the use of ribs. With a demagnification factor of 357, 2.2 mra d of the radiation from an NSLS bending magnet are focussed to a spot size with a FWHM of 363 mu m which is 12% larger than the ideal focus size with a beam intensity 2.0 X 10(9) photons s(-1) mA(-1) at 8 keV. The beam intensity changes by 5% when changing the sagittal bend radiu s from 1.81 to 4.09 m thereby demonstrating that the anticlastic curva ture effect is negligible.