COMPUTERIZED ANALYSIS OF LIF GR-200 TL SIGNALS - APPLICATION TO DOSE MEASUREMENTS IN THE MU-GY RANGE

Citation
A. Delgado et al., COMPUTERIZED ANALYSIS OF LIF GR-200 TL SIGNALS - APPLICATION TO DOSE MEASUREMENTS IN THE MU-GY RANGE, Radiation protection dosimetry, 60(2), 1995, pp. 147-153
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
ISSN journal
01448420
Volume
60
Issue
2
Year of publication
1995
Pages
147 - 153
Database
ISI
SICI code
0144-8420(1995)60:2<147:CAOLGT>2.0.ZU;2-L
Abstract
The features of a simple program for the computerised analysis of the LiF GR-200 glow curve are described. The adverse influence of the high residual signals characteristic of this material has been effectively eliminated for doses in the range 0.5-300 mu Gy. This program for TL evaluation, together with some improvements in the reader temperature measurement and control, permit one to obtain very good dose results i n this dose range employing only reader anneal. The good dosimetric pe rformance achieved under these simple working conditions has been show n in a detailed experiment whose results are also presented here. Very reproducible ambient dose results have been obtained in real working conditions exposing GR-200 dosemeters for times ranging between five h ours and three months.