STUDIES ON NI-63 BASED CHECK-LIGHT SOURCES FOR TL READERS

Citation
Js. Nagpal et al., STUDIES ON NI-63 BASED CHECK-LIGHT SOURCES FOR TL READERS, Radiation protection dosimetry, 60(2), 1995, pp. 181-184
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
ISSN journal
01448420
Volume
60
Issue
2
Year of publication
1995
Pages
181 - 184
Database
ISI
SICI code
0144-8420(1995)60:2<181:SONBCS>2.0.ZU;2-W
Abstract
The performance of indigenously produced, electrodeposited Ni-63 sourc es having 10 mm x 10 mm active area and individually coupled to a plas tic scintillator has been studied. TL phosphors such as CaF2, CaF2:Dy, CaF2:Tm, CaF2:Eu, TLD-100 and CaSO4:Dy with Ni-63 have been studied f or their possible use as check-light sources for testing TL readers. S tability at ambient temperatures (298-323 K), build-up and decrease of light output with time have been examined. A preliminary study of the fluorescent species in the gamma irradiated plastic scintillator has been made.