I. Abdelhamid et al., NEEDLE-TYPE GLUCOSE BIOSENSOR WITH AN ELECTROCHEMICALLY CODEPOSITED ENZYME IN A PLATINUM BLACK MATRIX, Electroanalysis, 7(8), 1995, pp. 738-741
In this study we report the development of a needle-type glucose biose
nsor for the management of hemorrhagic shock. The immobilization techn
ique depends on the electrophoretic deposition of the glucose oxidase
enzyme in an electrochemically grown platinum black matrix. The sensor
was coated with Nafion to decrease the effect of interferents. Prelim
inary tests were carried out to evaluate the sensor performance in vit
ro. These tests included the measurement of glucose levels in buffer s
olutions containing various potential physiological interferents, as w
ell as in bovine serum. The preliminary results show the sensor to hav
e linearity up to 33 mM and a diminished response to interferents. The
advantages of this technique are its simplicity and high controllabil
ity.