MULTI-LINE GAMMA-RAY SPECTROMETER PERFORMANCE OF A SI(LI) DETECTOR STACK

Citation
Gs. Hubbard et al., MULTI-LINE GAMMA-RAY SPECTROMETER PERFORMANCE OF A SI(LI) DETECTOR STACK, IEEE transactions on nuclear science, 42(4), 1995, pp. 441-444
Citations number
3
Categorie Soggetti
Nuclear Sciences & Tecnology","Engineering, Eletrical & Electronic
ISSN journal
00189499
Volume
42
Issue
4
Year of publication
1995
Part
1
Pages
441 - 444
Database
ISI
SICI code
0018-9499(1995)42:4<441:MGSPOA>2.0.ZU;2-T
Abstract
Experimental data is presented which for the first time displays multi -line spectrometer performance of a Si(Li) detector stack at elevated temperature. The stack consists of four elements, each with a 2 cm dia meter active area. Ba-133 and Ag-110m spectra are obtained using vario us techniques to enhance the peak-to-background ratio. Spectral data a re shown as a function of temperature (94 K less than or equal to T le ss than or equal to 230 K) using optimized peak shaping.