G. Baldazzi et al., A RADIATION DETECTION SYSTEM FOR HIGH-ENERGY COMPUTERIZED-TOMOGRAPHY USING CDZNTE DETECTORS, IEEE transactions on nuclear science, 42(4), 1995, pp. 575-579
This work presents a particular application in which cadmium-zinc tell
uride (CdZnTe) detectors are used in photoconductive mode in a high-en
ergy computerized tomography system for industrial applications. A fir
st-generation tomography facility has been set up at the University of
Bologna, based on a 12-MeV pulsed linear accelerator X-ray source and
a single 10x10x2 mm(3) CdZnTe detector.The obtained results show that
CdZnTe detectors exhibit better timing performances than CdTe detecto
rs, and are thus more suitable for tomographic applications. Special-p
urpose signal conditioning electronics have been developed in order to
compensate slow and fast signal fluctuations induced by temperature v
ariations and Linac instabilities. In such a way, the dynamic range of
the detector has been improved up to more than 3 orders of magnitude.
A set of images was acquired in order to evaluate the overall perform
ances of the tomographic facility.