Evr. Dibella et al., HETEROGENEITY OF SPECT BULLS-EYES IN NORMAL DOGS - COMPARISON OF ATTENUATION COMPENSATION ALGORITHMS, IEEE transactions on nuclear science, 42(4), 1995, pp. 1290-1296
In normal dogs, SPECT Tc-99m Sestamibi (MIBI) and Tl-201 myocardial pe
rfusion images reconstructed with filtered backprojection (FBP) show a
large decrease of counts in the septal wall (S) compared to the later
al wall (L). We evaluated the iterative method of Chang at 0 and 1 ite
rations (Chang0 and Chang1), and the Maximum Likelihood-Expectation Ma
ximization with attenuation compensation (ML-EM-ATN) algorithm on data
acquired from 5 normal dogs and from simulated projection data using
a homogeneous count-density model of a normal canine myocardium in the
attenuation field measured in one dog. Mean counts in the S and L reg
ions were calculated from maximum-count circumferential profile arrays
. The table shows mean +/- s.d. S/L count ratios. [GRAPHICS] Our resul
ts demonstrate that ML-EM-ATN and Chang1 result in improved uniformity
, as measured by the S/L ratio.