M. Kuriyan et A. Opalka, LEUKOREDUCED PLATELET APHERESIS PRODUCTION WITH A MODIFIED CODE SPECTRA COLLECTION PROTOCOL, Journal of clinical apheresis, 10(2), 1995, pp. 85-86
A platelet apheresis product or single donor platelet (SDP) with a whi
te blood cell content of 5 x 10(6) or below is considered optimal for
transfusion. We modified the COBE spectra platelet collection protocol
as follows: increased AC infusion rate of 1.1 ml/min/LTBV, setting th
e inlet:AC ratio configuration at high, and maintained the inlet flow
rate of this setting with a maximum of 80 ml/min. The collection time
varied with single dose or double dose production. The aim was to cons
istently obtain a platelet yield above 3 x 10(11) and a WBC content le
ss than 5 x 10(6). The results were tabulated on 64 donations. The pla
telet yield was >3 x 10(11) for 59 donations with a mean of 4.89, medi
an of 4.72, and a range of 3.02 to 8.14; and <3 x 10(11) for 5 donatio
ns with a mean of 2.74, median of 2.84, and a range of 2.14 to 2.96. T
he WBC content was <5 x 10(6) for 60 donations with a mean of 0.52, me
dian of 0.20, and a range of 0.06 to 0.71; and >5 x 106 for 4 donation
s with a mean of 59.9, a median of 48.9, and a range of 8.12 to 62. We
conclude that 93.8% of the donations yielded a SDP that met or exceed
ed leukoreduction standards. (C) 1995 Wiley-Liss, Inc.