EVALUATION OF THE SINGLE NEEDLE PROCEDURE IN PLATELETPHERESIS WITH THE FRESENIUS AS-104 BLOOD-CELL SEPARATOR

Authors
Citation
R. Moog et N. Muller, EVALUATION OF THE SINGLE NEEDLE PROCEDURE IN PLATELETPHERESIS WITH THE FRESENIUS AS-104 BLOOD-CELL SEPARATOR, Journal of clinical apheresis, 10(2), 1995, pp. 90-95
Citations number
40
Categorie Soggetti
Hematology
ISSN journal
07332459
Volume
10
Issue
2
Year of publication
1995
Pages
90 - 95
Database
ISI
SICI code
0733-2459(1995)10:2<90:EOTSNP>2.0.ZU;2-U
Abstract
The aim of the present study was to evaluate the single needle (SN) pr ocedure for plateletpheresis with the AS 104 blood cell separator. One hundred fifty-seven platelet collections with different protocols wer e carried out and analysed with respect to platelet yield, separation efficiency, leukocyte contamination, anticoagulant infused, and donati on time. SN plateletpheresis resulted in a mean platelet yield of 2.6- 3.1 x 10(11) with separation efficiencies of 49.2-57.4%. The mean leuk ocyte contamination ranged from 2.8 to 40.5 x 10(6). Parameters of pro duct quality were investigated in 14 platelet concentrates (PC). Deter mination of glucose, lactate, lactate dehydrogenase (LDH), and morphol ogy score showed good quality of the PC. (C) 1995 Wiley-Liss, Inc.