SENSITIVE DETERMINATION OF MANGANESE USING FLOW-INJECTION AND CHEMILUMINESCENT DETECTION

Citation
Ar. Bowie et al., SENSITIVE DETERMINATION OF MANGANESE USING FLOW-INJECTION AND CHEMILUMINESCENT DETECTION, Analyst, 120(8), 1995, pp. 2119-2127
Citations number
42
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
00032654
Volume
120
Issue
8
Year of publication
1995
Pages
2119 - 2127
Database
ISI
SICI code
0003-2654(1995)120:8<2119:SDOMUF>2.0.ZU;2-M
Abstract
The determination of Mn2+ in aqueous solution using a flow injection s ystem and chemiluminescent method of detection is reported. The Mn2+ i on is determined with high sensitivity and; selectivity by exploiting the Mn2+ catalysis of the luminous oxidation of 7,7,8,8-tetracyanoquin odimethane by dissolved O-2 in alkaline solution. This reaction is sen sitized by Eosin Y, arid further enhanced by means of didodecyldimethy lammonium bromide bilayer vesicles, The limit of detection for this re agent scheme is 4.5 ppb. A reaction mechanism and energy level diagram for this system is also presented for the first time. The method is a pplied to the determination of Mn2+ in raw and processed water.