CALCULATION OF SIGNAL INTENSITIES IN HYBRID SEQUENCES FOR FAST NMR IMAGING

Citation
R. Deichmann et al., CALCULATION OF SIGNAL INTENSITIES IN HYBRID SEQUENCES FOR FAST NMR IMAGING, Magnetic resonance in medicine, 34(3), 1995, pp. 481-489
Citations number
6
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
07403194
Volume
34
Issue
3
Year of publication
1995
Pages
481 - 489
Database
ISI
SICI code
0740-3194(1995)34:3<481:COSIIH>2.0.ZU;2-0
Abstract
A number of techniques that recently have been used for fast NMR-imagi ng are based on a hybrid sequence of echo planar imaging (EPI) and FLA SH imaging: after each NMR excitation several k-space lines are measur ed, The complete k-space is covered by performance of several excitati ons, It has been observed that there is usually an optimal hybrid sequ ence that maximizes the signal-to-noise ratio, In this work, a method is presented that allows a determination of the optimal sequence as a function of the relaxation times T-1 and T-2.