Pure-tone Reference Equivalent Threshold Sound Pressure Level (RETSPL)
of the ipsilateral stimulus receiver for acoustic reflex measurements
on Madsen Electronics type Zodiac 901 impedance audiometer is provide
d. The results, obtained from 20 normal-hearing subjects, are achieved
by comparing hearing threshold levels measured using a TDH 39 telepho
ne (calibrated to ISO 389) with thresholds recorded using the ipsilate
ral stimulus insert phone. The calibration is referenced to an IEC-711
ear simulator and comprises the following frequencies: 125, 250, 500,
750, 1000, 1500, 2000, 3000, 4000, 6000, 8000 Hz.