CALIBRATION OF IPSILATERAL STIMULUS TRANSDUCER FOR ACOUSTIC REFLEX MEASUREMENTS

Citation
So. Olsen et al., CALIBRATION OF IPSILATERAL STIMULUS TRANSDUCER FOR ACOUSTIC REFLEX MEASUREMENTS, Scandinavian audiology, 24(3), 1995, pp. 161-164
Citations number
7
Categorie Soggetti
Otorhinolaryngology
Journal title
ISSN journal
01050397
Volume
24
Issue
3
Year of publication
1995
Pages
161 - 164
Database
ISI
SICI code
0105-0397(1995)24:3<161:COISTF>2.0.ZU;2-Z
Abstract
Pure-tone Reference Equivalent Threshold Sound Pressure Level (RETSPL) of the ipsilateral stimulus receiver for acoustic reflex measurements on Madsen Electronics type Zodiac 901 impedance audiometer is provide d. The results, obtained from 20 normal-hearing subjects, are achieved by comparing hearing threshold levels measured using a TDH 39 telepho ne (calibrated to ISO 389) with thresholds recorded using the ipsilate ral stimulus insert phone. The calibration is referenced to an IEC-711 ear simulator and comprises the following frequencies: 125, 250, 500, 750, 1000, 1500, 2000, 3000, 4000, 6000, 8000 Hz.