V. Martindaguet et al., PACKED-COLUMN SUPERCRITICAL-FLUID CHROMATOGRAPHY - QUANTITATIVE-DETERMINATION OF URANIUM WITHOUT LIQUID WASTE GENERATION, Analytical chemistry, 69(3), 1997, pp. 536-541
A new procedure for the determination of uranium by packed-column supe
rcritical fluid chromatography is proposed. A nonfluorinated chelating
agent selective for copper and uranium, the 2,6-diacetylpyridin bis(b
enzoylhydrazone), has been chosen, We have studied its chromatographic
properties on different stationary phases and the influence of the me
thanol content in the carbon dioxide mobile phase, The separation of t
he metal compounds was conducted with and without solvent injection, A
calibration curve was obtained for uranium in the range of 52-323 ng
injected, The accuracy of the method is 0.5%,;the repeatability 4%. Th
e same studies were performed with a new compound, diacetyl-2,6 pyridi
ne bis(4-tert-butyl benzoylhydrazone). An increase in retention and ef
ficiency was then observed.