R. Kataky et al., POTENTIOMETRIC, ENANTIOSELECTIVE SENSORS FOR ALKYL AND ARYL AMMONIUM-IONS OF PHARMACEUTICAL SIGNIFICANCE, BASED ON LIPOPHILIC CYCLODEXTRINS, Scandinavian journal of clinical & laboratory investigation, 55(5), 1995, pp. 409-419
Lipophilic alpha, beta and gamma cyclodextrins (CDs) have been systema
tically analysed in order to establish their selective binding to oniu
m ions. They have been identified as excellent ionophores for a range
of ions of vital importance in clinical, pharmaceutical and forensic a
nalysis. From the results obtained we found the following. A peroctyla
ted cyclodextrin provides a size-selective cavity for the binding and
detection of onium ions. Potentiometric alpha, beta and gamma cyclodex
trin-based electrodes selectively sense NH4+, NMe(4)(+) and NEt(4)(+)
ions respectively. These experiments were performed in order to enable
us to understand the nature of analyte recognition by the lipophilic
cyclodextrins. Peroctylated alpha CD is a suitable ionophore for dopam
ine hydrochloride, - log [C] = 5.4, - log K = 2.0 (in serum level of N
a+, K+, Ca2+) whereas peroctylated and 2,6 didodecyl beta CD sense ace
tylcholine chloride, - log[C] = 5.0, - log K = 4.2 (in serum level of
Na+, K+, Ca2+) and creatinine hydrochloride, - log K = 2.7 (in serum l
evel of Na+, K+, Ca2+), respectively. The 2,6 didodecyl beta CD respon
ds to more bulky aryl ammonium ions such as the anaesthetics procaine,
prilocaine and lignocaine hydrochlorides, - log K = 4.2 (in serum lev
el of Na+, K+, Ca2+). Partially octylated alpha CD has been identified
as an enantioselective sensor for ephedrine hydrochloride, - log K =
4.5 (in serum level of Na+, K+, Ca2+), - log K-+/-(pot) = 2.6, and rel
ated compounds such as amphetamine hydrochloride. The 2,6 didodecyl be
ta CD is enantioselective for propranalol hydrochloride, - log K = 4.2
(in serum level of Na+, K+, Ca2+), - log K-+/-(pot) = 2.7. Complexati
on has also been studied by electrospray mass spectrometry (ESMS).