G. Reber et al., A NEW, SEMIQUANTITATIVE AND INDIVIDUAL ELISA FOR RAPID MEASUREMENT OFPLASMA D-DIMER IN PATIENTS SUSPECTED OF PULMONARY-EMBOLISM, Blood coagulation & fibrinolysis, 6(5), 1995, pp. 460-463
The performance of a new membrane ELISA for semi-quantitative determin
ation of plasma D-dimer has been evaluated. Its cut-off is about 500 n
g/ml FEU and this single test is completed within 10 min. D-dimer was
measured in 301 patients suspected of pulmonary embolism by convention
al microplate and membrane ELISA. For the latter, readings were made b
y eye and some differences were noticed between the readers for reacti
ons in the grey zone. Sensitivity and negative predictive values were
similar for the two ELISA (higher than 90%). The 95% confidence interv
als of sensitivity and negative predictive values obtained with this m
embrane ELISA suggest that this new test may be used as a diagnostic t
ool to exclude the presence of pulmonary embolism.